TY - JOUR UR - http://lib.ugent.be/catalog/pug01:2109243 ID - pug01:2109243 LA - eng TI - Functional atrial undersensing associated with switching to a tracking mode of pacing PY - 2012 JO - (2012) PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY SN - 0147-8389 PB - 2012 AU - Barold, S Serge AU - Van Heuverswyn, Frederic UZGent AU - Timmers, Liesbeth UZGent AU - Stroobandt, Roland GE35 802000891022 0000-0002-2835-5845 AB - We have previously demonstrated that contemporary St. Jude devices (pacemakers and implantable cardioverter-defibrillators [ICDs]; St. Jude Medical, Sylmar, CA, USA) are designed to generate an extended postventricular atrial refractory period (PVARP) of 475 ms at the termination of conventional automatic mode switching (AMS) in response to atrial tachyarrhythmias . This response may cause functional atrial undersensing . A similar PVARP response unrelated to conventional AMS was found in four St. Jude devices (three ICDs and one pacemaker) whenever a nontracking pacing mode switched to a tracking DDD(R) mode. PVARP extension and functional atrial undersensing were observed when the VOO, VVI, and the DDI(R) modes (unrelated to conventional AMS) switched to the DDD(R) mode . In one patient the switch from the OOO mode (in the programmed noise reversion mode) to the DDD mode occurred after cessation of electromagnetic interference disturbing the ventricular channel. In this case PVARP extension was seen only in the corresponding markers because no P waves occurred coincidentally with the extended PVARP. The PVARP extension caused by a mode switch to the tracking function was designed to prevent sensing of a retrograde P wave on the first cycle of the reestablished tracking mode. The observed functional atrial undersensing is a normal manifestation of device function and must not be misinterpreted as a true atrial undersensing problem. ER -Download RIS file
00000nam^a2200301^i^4500 | |||
001 | 2109243 | ||
005 | 20180813141435.0 | ||
008 | 120519s2012------------------------eng-- | ||
022 | a 0147-8389 | ||
024 | a 000309449600009 2 wos | ||
024 | a 1854/LU-2109243 2 handle | ||
024 | a 10.1111/j.1540-8159.2012.03461.x 2 doi | ||
040 | a UGent | ||
245 | a Functional atrial undersensing associated with switching to a tracking mode of pacing | ||
260 | c 2012 | ||
520 | a We have previously demonstrated that contemporary St. Jude devices (pacemakers and implantable cardioverter-defibrillators [ICDs]; St. Jude Medical, Sylmar, CA, USA) are designed to generate an extended postventricular atrial refractory period (PVARP) of 475 ms at the termination of conventional automatic mode switching (AMS) in response to atrial tachyarrhythmias . This response may cause functional atrial undersensing . A similar PVARP response unrelated to conventional AMS was found in four St. Jude devices (three ICDs and one pacemaker) whenever a nontracking pacing mode switched to a tracking DDD(R) mode. PVARP extension and functional atrial undersensing were observed when the VOO, VVI, and the DDI(R) modes (unrelated to conventional AMS) switched to the DDD(R) mode . In one patient the switch from the OOO mode (in the programmed noise reversion mode) to the DDD mode occurred after cessation of electromagnetic interference disturbing the ventricular channel. In this case PVARP extension was seen only in the corresponding markers because no P waves occurred coincidentally with the extended PVARP. The PVARP extension caused by a mode switch to the tracking function was designed to prevent sensing of a retrograde P wave on the first cycle of the reestablished tracking mode. The observed functional atrial undersensing is a normal manifestation of device function and must not be misinterpreted as a true atrial undersensing problem. | ||
598 | a A1 | ||
700 | a Barold, S Serge | ||
700 | a Van Heuverswyn, Frederic u UZGent 0 870110443455 9 FF0A54C6-F0ED-11E1-A9DE-61C894A0A6B4 | ||
700 | a Timmers, Liesbeth u UZGent 0 870110739206 9 0033F7D0-F0EE-11E1-A9DE-61C894A0A6B4 | ||
700 | a Stroobandt, Roland u GE35 0 802000891022 0 0000-0002-2835-5845 9 2E0B168E-F0EE-11E1-A9DE-61C894A0A6B4 | ||
650 | a Medicine and Health Sciences | ||
653 | a Cardiac pacing | ||
653 | a automatic mode switching | ||
653 | a electromagnetic interference | ||
653 | a atrial undersensing | ||
653 | a postventricular atrial refractory period | ||
773 | t PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY g PACE-Pacing Clin. Electrophysiol. 2012. 35 (10) p.1188-1193 q 35:10<1188 | ||
856 | 3 Full Text u https://biblio.ugent.be/publication/2109243/file/3090646 z [ugent] y PACE_2012_Funct_atrial_undersensing_AMS_tracking_PUBL.pdf | ||
920 | a article | ||
Z30 | x GE 1 GE01 | ||
922 | a UGENT-GE |
All data below are available with an Open Data Commons Open Database License. You are free to copy, distribute and use the database; to produce works from the database; to modify, transform and build upon the database. As long as you attribute the data sets to the source, publish your adapted database with ODbL license, and keep the dataset open (don't use technical measures such as DRM to restrict access to the database).
The datasets are also available as weekly exports.