TY - JOUR UR - http://lib.ugent.be/catalog/pug01:1984666 ID - pug01:1984666 LA - eng TI - Precision analysis of kinetic modelling estimates in dynamic contrast enhanced MRI PY - 2011 JO - (2011) MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE SN - 0968-5243 PB - 2011 AU - De Naeyer, Dieter UGent 002002144159 AU - De Deene, Yves AU - Ceelen, Wim GE38 UZGent 001985299303 802000604567 0000-0001-7692-4419 AU - Segers, Patrick TW06 801000864165 0000-0003-3870-3409 AU - Verdonck, Pascal AB - Dynamic contrast enhanced MRI and pharmacokinetic modelling provide a powerful tool for tumour diagnosis and treatment evaluation. However, several studies show low reproducibility of the technique and poor precision of the transendothelial transfer constant K (trans). This work proposes a theoretical framework describing how finite signal-noise-ratio (SNR) in the MR images is propagated throughout the measurement protocol to uncertainty on the kinetic parameter estimates. After deriving a distribution for the contrast agent concentration, a maximum likelihood estimator (MLM) is proposed that exhibits Cramer-Rao lower bounds (CRLB). An analytical expression is derived for the CRLB that can be used to determine confidence intervals for kinetic parameters and to investigate the influence of protocol parameters as scan time and temporal resolution on K (trans)-precision. K (trans)-uncertainty can be reduced up to 30% by using MLM in comparison with least square estimator. K (trans)-precision is proportional to the SNR and depends strongly on the kinetic parameter values themselves. Minimal scan time and temporal resolution were found to be 15 min and 15 s, respectively, for Gd-DTPA. Temporal resolution should be enhanced by decreasing the NEX parameter (NEX a parts per thousand currency sign 1). CRLB provide a golden standard to construct 95% confidence intervals, which can be used to perform protocol optimization and to test the statistical significance of K (trans)-changes in treatment evaluation. ER -Download RIS file
00000nam^a2200301^i^4500 | |||
001 | 1984666 | ||
005 | 20180813141258.0 | ||
008 | 120112s2011------------------------eng-- | ||
022 | a 0968-5243 | ||
024 | a 000288759200001 2 wos | ||
024 | a 1854/LU-1984666 2 handle | ||
024 | a 10.1007/s10334-010-0235-6 2 doi | ||
040 | a UGent | ||
245 | a Precision analysis of kinetic modelling estimates in dynamic contrast enhanced MRI | ||
260 | c 2011 | ||
520 | a Dynamic contrast enhanced MRI and pharmacokinetic modelling provide a powerful tool for tumour diagnosis and treatment evaluation. However, several studies show low reproducibility of the technique and poor precision of the transendothelial transfer constant K (trans). This work proposes a theoretical framework describing how finite signal-noise-ratio (SNR) in the MR images is propagated throughout the measurement protocol to uncertainty on the kinetic parameter estimates. After deriving a distribution for the contrast agent concentration, a maximum likelihood estimator (MLM) is proposed that exhibits Cramer-Rao lower bounds (CRLB). An analytical expression is derived for the CRLB that can be used to determine confidence intervals for kinetic parameters and to investigate the influence of protocol parameters as scan time and temporal resolution on K (trans)-precision. K (trans)-uncertainty can be reduced up to 30% by using MLM in comparison with least square estimator. K (trans)-precision is proportional to the SNR and depends strongly on the kinetic parameter values themselves. Minimal scan time and temporal resolution were found to be 15 min and 15 s, respectively, for Gd-DTPA. Temporal resolution should be enhanced by decreasing the NEX parameter (NEX a parts per thousand currency sign 1). CRLB provide a golden standard to construct 95% confidence intervals, which can be used to perform protocol optimization and to test the statistical significance of K (trans)-changes in treatment evaluation. | ||
598 | a A1 | ||
700 | a De Naeyer, Dieter u UGent 0 002002144159 0 802000264360 9 F98A47F4-F0ED-11E1-A9DE-61C894A0A6B4 | ||
700 | a De Deene, Yves u GE38 0 801000897511 9 F4B04670-F0ED-11E1-A9DE-61C894A0A6B4 | ||
700 | a Ceelen, Wim u GE38 u UZGent 0 001985299303 0 802000604567 0 0000-0001-7692-4419 9 FE50E55E-F0ED-11E1-A9DE-61C894A0A6B4 | ||
700 | a Segers, Patrick u TW06 0 801000864165 0 0000-0003-3870-3409 9 F4A2EB1A-F0ED-11E1-A9DE-61C894A0A6B4 | ||
700 | a Verdonck, Pascal u TW06 0 801000649755 9 F446FF9E-F0ED-11E1-A9DE-61C894A0A6B4 | ||
650 | a Technology and Engineering | ||
653 | a Experiment design | ||
653 | a Precision analysis | ||
653 | a Confidence interval | ||
653 | a TRACER KINETICS | ||
653 | a PHARMACOKINETIC PARAMETERS | ||
653 | a FLIP ANGLES | ||
653 | a DCE-MRI | ||
653 | a AGENT CONCENTRATION | ||
653 | a SIGNAL INTENSITY | ||
653 | a T-1-WEIGHTED MRI | ||
653 | a REPRODUCIBILITY | ||
653 | a UNCERTAINTY | ||
653 | a PERFUSION | ||
653 | a Kinetic modelling | ||
773 | t MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE g Magn. Reson. Mat. Phys. Biol. Med. 2011. 24 (2) p.51-66 q 24:2<51 | ||
856 | 3 Full Text u https://biblio.ugent.be/publication/1984666/file/1984753 z [ugent] y MagnResMatPhys_D._De_Naeyer.pdf | ||
920 | a article | ||
Z30 | x GE 1 GE10 | ||
922 | a UGENT-GE | ||
Z30 | x EA 1 TW15 | ||
922 | a UGENT-EA | ||
Z30 | x GE 1 GE17 | ||
922 | a UGENT-GE |
All data below are available with an Open Data Commons Open Database License. You are free to copy, distribute and use the database; to produce works from the database; to modify, transform and build upon the database. As long as you attribute the data sets to the source, publish your adapted database with ODbL license, and keep the dataset open (don't use technical measures such as DRM to restrict access to the database).
The datasets are also available as weekly exports.